The CPM-766 is an advanced Charged Plate Monitor that utilizes a microprocessor for assessing the effectiveness of air ionizers in neutralizing static charge within ionization systems. It offers the capability to conduct positive and negative decay tests, as well as balance (offset voltage) tests, enabling the determination of the operational efficiency of an ionization system. Compliant with the ESD Association Standard ANSI/ESD STM3.1 and IEC 61340-4-7 Ionization, the CPM-766 is suitable for testing various types of ionization systems.
The CPM-766 conducts both manual and automatic decay and balance tests to qualify and periodically verify ionizers. Its internal memory is capable of storing up to 200 test data. The test data includes balance averages, temperature and humidity, date and time and can be saved under specified locations and areas.
The CPM-766 is equipped with a Li-ion battery pack that can be recharged through a USB connection. With a battery life of over 250 hours when in standby, it enables you to conduct around 1000 tests on average.
When operating in Decay mode, an internal high voltage generator is utilized to charge the plate to a voltage up of 1250 volts. During the test, the CPM-766 observes the duration required for the conductive plate to decay from 1000 volts to the stopping voltage selected by the user.
In the Balance mode, the Charged Plate Monitor will measure the offset voltage on the floating plate to determine an imbalance of positive and negative ions of an ionizer. An offset voltage threshold can be set to 35, 100 volts or 200 volts, where the information on the display during the test turns red to notify the user.
The CPM-766 provides the ability to perform automated Decay tests that can include a Balance test, in sequence. The number of Decay cycles and Decay sequences are determined by the user.
In Manual Operation you will be able to isolate the plate, charge the plate up to ±1,250 volts, ground the plate and control the timer for custom testing.
Using a large 5” color touchscreen LCD display, the CPM-766 displays clear, easy-to-use controls and a full keyboard that allows customization of labels, even while wearing gloves.
The CPM-766 is equipped with 2 standard ground jacks, an analog output jack, and a single USB-C port that provides connectivity to your PC for software updates and charging.
The 6” x 6” isolated plate is not detachable and cannot be removed.
CPM-766 Charged Plate Monitor Specifications
Performance |
|
Charging Range |
±1,250 ±5% volts DC |
Charging Speed |
Charges from zero to over ±1000 volts in 3 seconds at ambient conditions |
Accuracy |
Voltage Monitor Output: Better than ±5% of reading, ±10 mV
Voltage Display: Better than ±5% of reading, ±2 counts
|
Response Time |
150ms (90-10%)
|
3dB Bandwidth |
6.5 Hz
|
Sensor Noise Signal (Typical)1 |
Maximum Voltage: +3 volts
Minimum Voltage: -5 volts
|
Output |
|
Connection Type |
BNC Connector located on the back panelt |
Analog Output |
±2 volts |
Scale Factor |
1 volt output corresponds to 10 kV |
Isolated Conductive Plate |
|
Conductive Plate Size |
6.0" x 6.0"
15.2 cm x 15.2 cm
|
Capacitance |
20 pF ±2 pF |
Plate Voltage Retention |
Floating plate voltage discharges less than 10% over a 5 minute period when charged to over 1kV2 |
Plate Spacing |
0.805” (2.4 cm) between isolated and ground plates |
Spacers |
Plate spacers made from machined virgin, white Teflon® |
Plate Construction |
Aluminum
Bright nickel plating
Conductive plate is not detachable
|
Display |
|
Type |
Color Capacitive Touchscreen |
Size |
5.0” Diagonal |
Resolution |
800 x 480 pixels |
Response Time |
10ms Typical |
Surface Treatment |
Anti-Glare |
Touch Point |
5 |
Glove Use |
Supports PVC, PE, Lightweight Rubber Gloves (0.3mm thick) |
Grounding |
|
Case |
2 green banana jacks mounted on the side panel |
Temperature & Relative Humidity Sensor |
|
Temperature Accuracy3 |
±2°F (±1°C) |
Relative Humidity Accuracy |
±10% |
Power |
|
Battery |
Re-chargeable Li-Ion battery (included). Battery swappable through easily accessible battery door at the bottom of the instrument |
Battery Life |
250 hours when in standby mode
48 hours typical
10 hours (of continuous use)
Approximately 1000 decay tests
|
Charging |
Rechargeable via USB 2.0 through PC port or Wall Charger (included) |
Charging Time |
4 hours (from 1 to 100%) |
Memory |
|
Capacity |
200 Data Sets |
Data Sets |
Data sets include Decay Time, Start and Stop Voltage, Min/Max/Avg, Temperature & Humidity, Location, Area and Timestamp |
Environment |
|
Operating Temperature |
10°C to 30°C (50°F to 80°F) |
Operating Relative Humidity |
0% – 80%, non-condensing |
Operating Altitude |
2000 m |
Physical Specifications |
|
Dimensions (HxWxL)4 |
5.5” x 8.9” x 10.6”
13.9 cm x 22.6 cm x 26.9 cm
|
Weight |
4.83 lbs (77.28 oz.) with battery
4.69 lbs (75.04 oz.) without battery
|
Case Material |
Aluminum
|
Warranty |
|
|
2 years on circuit board
1 year on the display
1 year on sensor5 |
1 Typical specifications are not guaranteed.
2 When measuring at <70% Rh
3 Accuracy based on the instrument while not actively charging
4 When in overhead measurements position
5 Warranty on the sensor limited to a defective sensor that was not dropped or used for measuring a source greater than 20kV.
801B-012 Li-Ion Battery Pack
General Specifications |
|
Battery Type |
3-cell, Rechargeable |
Technology |
Lithium-Ion Technology |
Nominal Capacity |
2400mAh |
Output Voltage |
3.7V |
Performance Amp-Hour |
2.4Ah |
Performance Watt-Hour |
8.88Wh |
Transport Safety Certified |
UN38.3 |
IATA UN Number |
UN3480 |
IATA Class (Sub Hazard) |
9 |
Operating Temperature |
Discharging: -20°C to 60°C (-4°F to 140°F)
Charging: 0°C to 45°C (32°F to 113 °F) |
Storage Temperature |
-5°C to 35°C (-23°F to 95°F) |
Storage Humidity |
≤75% Rh |
Battery Pack Dimensions (WxHxD) |
53mm x 48.5mm x 15mm
2.0” x 1.9” x 0.6” |
Weight |
65g (2.2 oz) |
Packing Instructions (P.I.)1
When battery is packaged separately (e.g. a replacement battery pack):
|
IATA P.I. 965 Section IB applies |
When battery is packaged with the resistance meter, not contained in it: |
IATA P.I. 966 Section II applies |
When battery is contained within the resistance meter, then packaged: |
IATA P.I. 967 Section II applies |
1Per IATA 2021 regulations. Regulations subject to change without notice.
The 801B-012 battery packs have been tested and were found to comply with the criteria of “UN Model Regulations, Manual of Test and Criteria, ST/SG/AC.10/11/Rev.7 Part III, subsection 38.3”, also known as “UN38.3”. As a result they can be shipped unrestricted internationally by any means.
Ensure that any shipment packaging that contains these batteries is properly marked on the outside of the package for containing Li-ion batteries, using the label as described in the ‘Additional Requirements Section’ of Packing Instructions 965...970. Minimum size of the label is 120 x 110 mm (4.75 x 4.33 inches).